PRODUCT SPECIFICATIONS
Veterinary Infusion Pump IPA112
| Model No | IPA112 |
|---|---|
| Range of Flow Rate | 0.1-1200ml/h |
| Infusion Mode | Rate, Rate-Time, Rate-Volume, Time-Volume, / Drop-Volume, Drop-Time |
| Drug Library | 20 drug 11st with drug code display |
| Purge/Bolus Rate | 1-1200ml/h, defalut 800ml/h, step by 1ml/h |
| Single Bolus Volume | 1.0-10ml Adjustable, default 3 ml |
| Time Preset | 00: 01~99: 59 (Hour: Min) |
| Volume Range | 1-9999ml |
| Drop Range | 1-400d/min, step by 1 drop |
| Accuracy | ±5% |
| Total Volume Infused | 0-9999 ml |
| Air in line Detection | Ultrasonic wave |
| Visible & Audible Alarms | Air in line, Down Stream Occlusion, Door Open, VTBI Completion, / Near Completion, Low Battery, Battery Exhausted, Low Temperature, Motor Malfunction, AC Disconnection, AC connection, I.V set / dislocation, MPU Error, Batery Charing, Battery Charing Completion, Circuit Malfunction |
| K.V.O Rate | 1ml/h-5ml/h, default value 1ml/h, can be programmed by user step 0.1ml/h |
| Internal Battery | Lithium battery, 11.1/2000mAh, more than 4 hours working back up |
| Power Consumption | 30VA |
| Power | AC 100V-240V 50HZ/60HZ |
| Classication | Class II, Type CF, IPX4 |
| Dimension & Weight | 13x17.5x23 cm; 2Kg |
| Optional Function | Ambulance DC 12V |